HC
40XV CiTicel
delivery time: in stock
How it works: Electrochemical
Measuring range: 1-25%
Overload: 30%
Output signal: 0.10±0.02mA in air
Response time (T90): <15S
Zero current (bias): <0.6%
Linear: can be considered linear in many cases
Load resistance: 100 ohms
Shell material: ABS
Weight: <16 g
Orientation sensitivity: <0.2%
Operating temperature range: -20°C to +50°C
Storage temperature: 0℃-20℃
Thermal transient (temperature slump +22°C to -20°C): <23.5%
Working pressure range: ±20% of atmosphere
Pressure coefficient: <0.02%/Mbar
Pressure dynamics (60 cm H2O step change): <200% signal change
Continuous relative humidity range (at 0°C to 20°C): 5 to 95% RH non-condensing
Short-term relative humidity range (at 0°C to 20°C): 0 to 99% RH non-condensing
Long term output drift: <2%/month, typically <5% over operating life