HC
Tester, 1002-892, Megger S1-1568
delivery time:8-10 weeks
AC voltage input range
90-264V rms, 50/60 Hz, 200A
Battery
11.1V, 5.2Ah Complies with IEC 62133:2003 (S1-1568 has 2 batteries)
Battery Life
Typically 4.5 hours under 15kV, 100MΩ load
battery charging time
2.5 hours after deep discharge
2 hours, after normal discharge
30 minutes fast charge: 1 hour operation at 5kV, 100 MΩ
Test voltage
250V, 500V, 1000V, 2500V, 5000V, 10000V, 15000V, VL
Test voltage lockout
40V to 1kV in 10V steps
1kV to 5kV in 25V steps
5kV to 10kV in 25V steps
Voltage output accuracy
+4%, -0%, ±10V normal test voltage@1GΩ load (0℃~30℃)
Impedance range
10kΩ to 35TΩ
shielding performance
Shielding 500kΩ leakage resistance under 100MΩ load yields 1% test error
Analog display range
100kΩ to 10TΩ
Digital display range
10kΩ to 35TΩ
short circuit current
6mA standard
recharging current
2mA
Insulation alarm
100kΩ to 10GΩ
Capacitor rechargeable battery power
<2.5s/μF to 5kV, <5s/μF to 10kV
Capacitor charging AC power
<1.5s/μF to 5kV, <2.7s/μF to 10kV
capacitor discharge
5kV to 50V: <120ms/μF
10kV to 50V: <250ms/μF
15kV to 50V: <3500ms/μF
Capacitance range
10nF to 50µF (depending on measurement voltage >500V)
Capacitance Accuracy (23℃)
±10%±5nF
Current measurement range
0.01nA to 8mA
Accuracy (23℃)
±5%±0.2nA for all voltages
Interference suppression
8mA at 2800V to 15kV
software filtering
4 filter settings: 10s, 30s, 100s, 200s
Voltmeter range
30V to 660V AC or DC, 45Hz-65Hz
Voltmeter Accuracy
±3%, ±3V
timer range
Maximum 99 minutes 59 seconds, minimum 15 seconds
memory capacity
11 hour recording @ 5 second interval
testing method
IR, IR(t), DAR, PI, DD, SV, ramp test
interface
USB type B (device), Bluetooth®
real-time output
Reading/s(V, I, R) speed is 1Hz
remote control
Remote control can only be operated via USB cable (requires RC indicator in place)
Security Level
Complies with IEC 61010-1
CAT IV 1000V to 4000m (15 kV model)
Applicable height
4000m
operating temperature
-20℃ to 50℃
Storage temperature
-25℃ to 65℃
humidity
90%RH non-condensing@40℃
Ingress protection
IP 65 (lid closed), IP 40 (lid open)
safety
Meets IEC 61010-1, CAT IV 1000V requirements
EMC
Comply with IEC61326-1 requirements
volume
305mm×194mm×360mm (15kV model)
weight
6.5kg