HCM
9100S-B
Delivery time:6-8 weeks
Range | 95 to 707°F (35 to 375°C) |
Accuracy |
±0.25°C at 50°C ±0.25°C at 100°C ±0.5°C at 375°C |
Stability |
±0.07°C at 50°C ±0.1°C at 100°C ±0.3°C at 375°C |
Resolution | 0.1°C or °F |
Well-to-Well Uniformity | ±0.2°C with sensors of similar size at equal depths within wells |
Computer Interface | RS-232 included with Interface-It software |
NIST-Traceable Calibration | Data at 122°F (50°C), 212°F (100°C), 302°F (150°C), 392°F (200°C), 482°F (250°C), 572°F (300°C), and 707°F (375°C) |
Heating Times | 95 to 707°F (35 to 375°C): 9.5 minutes |
Stabilization | 5 minutes |
Cooling Times | 707 to 212°F (375 to 100°C): 14 minutes |
Well Depth |
4" (102 mm) 1⁄16" (1.6 mm) hole in 3.5" (89 mm) deep |
Power | 115 V AC (±10%), 55 to 65 Hz, 1.5 A or 230 V AC (±10%), 0.8 A, 45 to 55 Hz, 175 W |
Dimensions | 2.25 x 4.9 x 6.1" (57 x 125 x 150 mm) |